Edan M3 Vital Signs Monitor

$2,050.00 + GST

The EDAN M3 is a LCD based monitor that will fit well with your outpatients department or GP practice. A wide range of accessories avaiilable.
Please allow 2 – 3 weeks for delivery.


SKU EDAN-M3SPNI Categories , ,


Share
Description

Edan_M3_Vital_Signs_Brochure

EDAN_M3_Vital Signs_Monitor_User_Manual

Edan M3 Vital Signs Monitor

A super sleak update for the M3A with LCD screen morphs into the M3. Same reliability and added features on the new, including an upgrade path the Nellcor Oximax oximetre to help your site standardise.

Physical Specifications

Dimension: (174+2) mm (W)x (235z2) mm (H)x(189+2) mm(D)
Weight 3.5kg Excluding battery

 

Power Supply

Power Supply 100 V to 240 V-, 50 Hz/60 Hz
Pmax 70 VA

Battery
Battery Type :Rechargeable lithium-ion battery

 

Typical Working Time
2500 mAh 7 h
5000 mAh 14 h

Recharge Time
2500 mAh 3.5 h
5000 mAh 7 h

Display

Display screen: 5.6-inch color TFT
Resolution: 640X480

 

Data Storage

Trend: 72 h, at 1 min resolution
NIBP trend list: 30,000 groups
Alarm list: 800 groups

Recorder

Record Width: 58mm
Paper Speed 12.5mm/s. 25mm,s

Recording types:
Parameter list recording
Trend graph review recording
Alarm list recording
Real-time 8s waveform recording
Recording of all the data of current patient ID
Real-time alarm recording

Wireless Network:

Compliant with Standard and Directive: IEEE802.11b/g, R&TTE Directive (99/5/EEC)

Frequency Range
2.412 GHz -2.462 GHz (America)
2.412 GHz -2.484 GHz (Japan)
2.412 GHz -2.472 GHz (ETSI)

Interface and others:

USB Port: 1
Newrowrn Interface: 1
Nurse Call: RJ45 network interface

 

NIBP

Method: Oscillometric
Mode: Manual, Auto, Continuous,
Average Measuring Interval in Auto Mode: /2/3/4/5/10/15/30/60/90/120/240/480 min
Continuous: 5 min, interval is 5 s
Measuring Type: SYS, DIA, MAP, PR
Average measurement:
Interval: 1/2/3/4/5 min
Times: 3/5

Measuring Range

Adult Mode:
SYS: 40 mmHg to 270 mmHg Adult Mode
DIA: 10 mmHg to 215 mmHg
MAP: 20 mmHg to 235 mmHg

Paediatric Mode:
SYS: 40 mmHg to 230 mmHg Pediatric Mode
DIA: 10 mmHg to 180 mmHg
MAP: 20 mmHg to 195 mmHg

Neonate Mode:
SYS: 40 mmHg to 135 mmHg Neonatal Mode
DIA: 10 mmHg to 100 mmHg
MAP: 20 mmHg to 110 mmHg

Cuff Pressure Measuring Range: 0mmHg to 300 mmHg
Pressure Resolution: 1mmHg
Maximum Mean Error +/-5 mmHg
Maximum Standard Deviation: 8 mmHg
Maximum Measuring Period:
Adult/ Pediatric          120 s
Neonatal         90 s
Typical Measuring Period: 20 s to 35 s (depend on HR/motion disturbance)

Overpressure Protection
Adult: 297 mmHg +/-3 mmHg
Pediatric: 245 mmHg +/-3 mmHg
Neonatal 147 mmHg +/-3 mmHg

PR

Measuring range: 40 bpm to 240 bpm
Accuracy: +/-3 bpm or 3.5%, whichever is greater

SpO2

EDAN Module
Measuring Range: 0% to 100%
Resolution: 1%
Data update period: 1 s

Accuracy:
Adult/Pediatric: +/-2% (70% to 100% SpO2 )
Undefined (00/o to 690/o SpO2 )

Neonatal:  +/-3% (70% to 100% SpO2 )
Undefined (0% to 69% SpO2 )

PI (Perfusion Index)
Measuring Range: 0-1
Ressolution: 1

Pulse Rate
Measuring Range 25 bpm to 300 bpm
Resolution: 1 bpm
Accuracy +/-2 bpm

Nellcor Module

Accuracy:
DS-100A, OXI-A/N(Adult) +/- 3% (70%-100% SpO2)
OXI-A/N(Neonate) +/- 4% (70%-100% SpO2 )

Pulse Rate:
Measuring Range 20 bpm to 300 bpm
Resolution: 1 bpm
Accuracy +/-3 bpm  (20 bpm to 250 bpm)

TEMP

T2 Module (EDAN Quick TEMP)
Monitor mode: 25°C -45°C
Predict mode: 35.5°C -42°C
Sensor type: Oral /Auxiliary/RectaI
Resolution: 0.1 O C
Accuracy: Monitor mode: +/-.1°C (25°C -45°C )
Response Time: < 60 s
Time for Predicting: < 30 s
Measuring Mode: Direct Mode/Adjusted Mode

TH Module (IR Ear Temp)
Measuring Range: 34~42.2 °C
Resolution: 0.1 °C
Response Time: 0.1s

Clinical Accuracy
+/-0.2°C (35.5~42°C )
+/-0.3°C (out of the range mentioned above)
Laboratory Accuracy: +/- 0.2°C

F3000 Module (Covidien Quick Temp)
Measuring Range: 30~43 °C
Prediction Measuring Range: 35~43 °C
Cold Mode Measuring Range: 35~43 °C
Sensor Tyoe: Oral /Auxiliary/RectaI
Resolution: 0.1 °C
Accuracy: Monitor mode Predictivr mode: +/- 0.1°C
Quick Predictive mode: +/- 0.3°C

Typical Measurement time:
Oral (Quick Predictive Mode): (3~5) s (non-fever temps); (8~10) s (fever temps)
Oral (Predictive Mode): (6 ~ 10) s
Axillary: (8 ~12) s
Rectal: (10 Z~14) s
Monitoring Mode (all sites): (60 –      120) s
Measuring Mode: Direct Mode /Adjusted Mode

 

Compliant with Standards

IEC 60601-1: 2005-›A1:2012;
EN 60601-1: 2006+A1: 2013; IEC 60601-1-2:2007; EN 60601-1-2: 2007; IEC 80601-2-30: 2009;
ISO 80601-2-61: 2011; ISO 80601-2-56: 2009; IEC 60601-2-49:2011

Additional information
Weight10 kg